Meter tab
- Last UpdatedNov 18, 2025
- 6 minute read

The table below describes the commonly selected fields on the Meter Properties: Meter Tab.
|
Section |
Field |
Description |
|---|---|---|
|
Meter |
||
|
Name |
The name of the meter record. Name can be up to 31 characters in length. |
|
|
Input Unit Profile |
Unit profile with the units of the incoming data. |
|
|
Output Unit Profile |
Unit profile with the units used for reporting and billing. |
|
|
Remote |
Remote Connectivity |
|
|
Meter General |
||
|
Metering Method |
Defines the meter type. Choices are the following:
|
|
|
Accumulator Meter |
Select the checkbox if your metering device is of the accumulator-type. For more information, refer to If your metering device gathers accumulator data. |
|
|
Accumulator Rollover |
Enter the maximum value that the accumulator goes up to before it resets. For more information about how Measurement accounts for the rollover, refer to Account for an accumulator's rollover. The rollover value entered here is assumed to be in the units of the value being calculated. |
|
|
Field device local timezone |
The time zone in which the field device collecting the metering data is located. |
|
|
Field device contract hour |
The contract hour of the field device. The Field Device Contract Hour can be used to timestamp the daily aggregated rows with the contract hour and time zone used by the field device. If CD_FIELD_DEVICE_AGGREGATES is set to Y in the configDesc table, the contract hour and time zone will be supplied by the field. |
|
|
Official Contract Hour |
The Official Contract Hour can be used to timestamp the daily aggregated rows with the contract hour used by the system's time zone, as opposed to using the contract hour and time zone supplied by the field. The system's time zone is set in the configDesc table, in CD_CONTRACT_TIMEZONE_CODE. If the customer wishes to use the Official Contract Hour according to the system time (using the time zone entered in CD_CONTRACT_TIMEZONE_CODE), as opposed to the information supplied by the field, they need to set CD_FIELD_DEVICE_AGGREGATES to N in the configDesc table. |
|
|
Alternate Contract Hour |
This allows the customer to supply an Alternate Contract Hour. Note: The Alternate Contract Hour is not used for any data processing. It is purely informational. |
|
|
Supercompressibility calculation |
Select your supercompressibility calculation method from the dropdown list:
|
|
|
Telemetry calculation policy |
Defines the recalculation policy for volume, energy, and hydrocarbon dewpoint (if hydrocarbon dewpoint module is installed) when reading the data. The choices are the following:
|
|
|
Static pressure setting |
Defines whether the pressure values received are absolute or gauge. |
|
|
Config calculation policy |
Defines the recalculation policy for volume, energy and volume ratio with respect to configuration changes. The choices are as follows: Calculate — Calculate volume/energy or volume ratio on a config change. None — Do not calculate volume/ energy or volume ratio on a config change. |
|
|
Enable flow editor calculations |
Determines whether energy and/or volume calculations are performed when you edit temperature or pressure in the Meter Data Editor window. |
|
|
Enable flow editor dewpoint calc |
Determines whether hydrocarbon dewpoint calculations are performed when volume recalculation is performed in the Meter Data Editor window. |
|
|
Calibration |
AVEVA Measurement Advisor will test for variances in volume and energy percentages and determine if corrections are needed. |
|
|
Percent Energy Threshold |
AVEVA Measurement Advisor calculates the difference between the corrected energy and the current energy. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value. |
|
|
Percent Volume Threshold |
AVEVA Measurement Advisor calculates the difference between the corrected volume and the current volume. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value. |
|
|
Absolute Energy Threshold |
AVEVA Measurement Advisor calculates the difference between the corrected energy and the current energy. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value. |
|
|
Absolute Volume Threshold |
AVEVA Measurement Advisor calculates the difference between the corrected volume and the current volume. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value. |
|
|
Meter History |
||
|
Primary history granularity |
Defines the granularity at which the data is received from the field device. The choices are as follows:
|
|
|
Meter Analysis |
||
|
ZQ calculation granularity |
The granularity of the zone gas analysis to use in calculations requiring zone gas quality (GQ). |
|
|
MQ calculation granularity |
The granularity of the measured gas analysis to use in calculations requiring measured GQ. |
|
|
Recalculation policy |
Defines the recalculation policy for volume, energy and hydrocarbon dewpoint (if the hydrocarbon dewpoint module is installed) when the gas analysis data changes. The choices are the following:
|
|
|
ZQ download delay (hours) |
When the ZQ (that is, a chromatograph) is upstream from the meter and the gas takes a few hours to reach this meter, you can delay the download of the ZQ to the meter by this amount. |
|
|
Backup GQ Recalculation Policy |
Defines the recalculation policy for volume, energy, and hydrocarbon dewpoint (if the hydrocarbon dewpoint module is installed) when reading the data. The choices are the following:
|
|
|
Host GQ calculation policy |
Defines the gas analysis data used when the meter is recalculated according to the recalculation policy. The choices are as follows:
|
|
|
MTR Missing Data Check Note: This property is only available if you have the Estimation module installed. |
||
|
Enable |
Enables missing data detection for this meter. |
|
|
Enable Linear Interpolation |
Use linear interpolation to estimate the missing data. |
|
|
MTR Late Data Check Note: This property is only available if you have the Estimation module installed. |
||
|
Enable |
Enables the check for late data for this meter. |
|
|
Meter Frequency |
Defines the frequency at which the data is expected to arrive. |
|
|
MTR Estimation Note: This property is only available if you have the Estimation module installed. |
||
|
Data estimation type |
Defines the type of estimation used when missing or late data is detected (if enabled). The choices are as follows:
|
|