Please ensure Javascript is enabled for purposes of website accessibility
Powered by Zoomin Software. For more details please contactZoomin

AVEVA™ Measurement Advisor

Meter tab

  • Last UpdatedNov 18, 2025
  • 6 minute read

The table below describes the commonly selected fields on the Meter Properties: Meter Tab.

Section

Field

Description

Meter

Name

The name of the meter record.

Name can be up to 31 characters in length.

Input Unit Profile

Unit profile with the units of the incoming data.

Output Unit Profile

Unit profile with the units used for reporting and billing.

Remote

Remote Connectivity

Meter General

Metering Method

Defines the meter type.

Choices are the following:

  • Orifice Meter — A meter measuring flow by orifice technology.

  • Pulse Meter — A meter measuring flow by pulse counting technology.

  • Coriolis Meter — A metering device that measures mass flow rate by measuring tube displacement resulting from Coriolis effect.

  • Manual Meter — A manual meter is used to collect information when there is no telemetry. It typically consists volume and energy information. But if it is associated with a Gas Quality record, the energy can be calculated from the Volume. Volume can be calculated from the mass for manual meters with a Data Source of Manual Meter.

  • Balance Calculated — When a station volume is calculated from several meter runs, the result can be stored as meter data instead of a station volume record. This allows for the system to apply the full inventory of meter validation operations.

  • Line Pack — A meter measuring volume between two points.

  • Corrector Meter — A volume correction measurement device.

Accumulator Meter

Select the checkbox if your metering device is of the accumulator-type. For more information, refer to If your metering device gathers accumulator data.

Accumulator Rollover

Enter the maximum value that the accumulator goes up to before it resets. For more information about how Measurement accounts for the rollover, refer to Account for an accumulator's rollover. The rollover value entered here is assumed to be in the units of the value being calculated.

Field device local timezone

The time zone in which the field device collecting the metering data is located.

Field device contract hour

The contract hour of the field device. The Field Device Contract Hour can be used to timestamp the daily aggregated rows with the contract hour and time zone used by the field device.

If CD_FIELD_DEVICE_AGGREGATES is set to Y in the configDesc table, the contract hour and time zone will be supplied by the field.

Official Contract Hour

The Official Contract Hour can be used to timestamp the daily aggregated rows with the contract hour used by the system's time zone, as opposed to using the contract hour and time zone supplied by the field.

The system's time zone is set in the configDesc table, in CD_CONTRACT_TIMEZONE_CODE.

If the customer wishes to use the Official Contract Hour according to the system time (using the time zone entered in CD_CONTRACT_TIMEZONE_CODE), as opposed to the information supplied by the field, they need to set CD_FIELD_DEVICE_AGGREGATES to N in the configDesc table.

Alternate Contract Hour

This allows the customer to supply an Alternate Contract Hour.

Note: The Alternate Contract Hour is not used for any data processing. It is purely informational.

Supercompressibility calculation

Select your supercompressibility calculation method from the dropdown list:

  • NX-19 GCN

  • NX-19 Complete

  • AGA-8 1992 Detail

  • AGA-8 1992 Gross 1

  • AGA-8 1992 Gross 2

  • AGA-8 2017 Detail

  • AGA-8 2017 Gross 0

  • AGA-8 2017 Gross 1

  • AGA-8 2017 Gross 2

  • AGA-8 2017 Gross 3

  • AGA-8 2017 GERG

    Notes: For the purposes of the Supercompressibility calculation, AGA-8 2017 Detail and AGA-8 2017 Gross 1 and 2 are the same as their AGA-8 1992 counterparts.

    The AGA-8 2017 supercompressibility options cannot be used in conjunction with host-calculated meters or gas qualities (GQ) at this time. This functionality is expected in a future release.

Telemetry calculation policy

Defines the recalculation policy for volume, energy, and hydrocarbon dewpoint (if hydrocarbon dewpoint module is installed) when reading the data. The choices are the following:

  • Recalculate Volume And Energy

  • Recalculate Energy

  • None

  • Recalculate Volume, Energy, and Dewpoint

  • Recalculate Energy and Dewpoint

  • Recalculate Dewpoint

Static pressure setting

Defines whether the pressure values received are absolute or gauge.

Config calculation policy

Defines the recalculation policy for volume, energy and volume ratio with respect to configuration changes. The choices are as follows:

Calculate — Calculate volume/energy or volume ratio on a config change.

None — Do not calculate volume/ energy or volume ratio on a config change.

Enable flow editor calculations

Determines whether energy and/or volume calculations are performed when you edit temperature or pressure in the Meter Data Editor window.

Enable flow editor dewpoint calc

Determines whether hydrocarbon dewpoint calculations are performed when volume recalculation is performed in the Meter Data Editor window.

Calibration

AVEVA Measurement Advisor will test for variances in volume and energy percentages and determine if corrections are needed.

Percent Energy Threshold

AVEVA Measurement Advisor calculates the difference between the corrected energy and the current energy. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value.

Percent Volume Threshold

AVEVA Measurement Advisor calculates the difference between the corrected volume and the current volume. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value.

Absolute Energy

Threshold

AVEVA Measurement Advisor calculates the difference between the corrected energy and the current energy. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value.

Absolute Volume

Threshold

AVEVA Measurement Advisor calculates the difference between the corrected volume and the current volume. If the difference is greater than the threshold, then corrections will be required. You can enter values either as a percentage or as an absolute value.

Meter History

Primary history granularity

Defines the granularity at which the data is received from the field device. The choices are as follows:

  • Hourly

  • Daily

  • 15-minute

  • Monthly

  • Irregular

Meter Analysis

ZQ calculation granularity

The granularity of the zone gas analysis to use in calculations requiring zone gas quality (GQ).

MQ calculation granularity

The granularity of the measured gas analysis to use in calculations requiring measured GQ.

Recalculation policy

Defines the recalculation policy for volume, energy and hydrocarbon dewpoint (if the hydrocarbon dewpoint module is installed) when the gas analysis data changes.

The choices are the following:

  • None

  • Recalculate Energy

  • Recalculate Volume and Energy

  • Recalculate Volume, Energy, Dewpoint

  • Recalculate Energy and Dewpoint

  • Recalculate Dewpoint

ZQ download delay (hours)

When the ZQ (that is, a chromatograph) is upstream from the meter and the gas takes a few hours to reach this meter, you can delay the download of the ZQ to the meter by this amount.

Backup GQ Recalculation Policy

Defines the recalculation policy for volume, energy, and hydrocarbon dewpoint (if the hydrocarbon dewpoint module is installed) when reading the data. The choices are the following:

  • Recalculate Volume And Energy

  • Recalculate Energy

  • None

  • Recalculate Volume, Energy, and Dewpoint

  • Recalculate Energy and Dewpoint

  • Recalculate Dewpoint

    Note: This property is only used if the GQ is using a backup GQ and if the Recalculation policy is set to None.

Host GQ calculation policy

Defines the gas analysis data used when the meter is recalculated according to the recalculation policy. The choices are as follows:

MTR Missing Data Check

Note: This property is only available if you have the Estimation module installed.

Enable

Enables missing data detection for this meter.

Enable Linear

Interpolation

Use linear interpolation to estimate the missing data.

MTR Late Data Check

Note: This property is only available if you have the Estimation module installed.

Enable

Enables the check for late data for this meter.

Meter Frequency

Defines the frequency at which the data is expected to arrive.

MTR Estimation

Note: This property is only available if you have the Estimation module installed.

Data estimation type

Defines the type of estimation used when missing or late data is detected (if enabled).

The choices are as follows:

  • MTR Estimation (Default)

  • MTR Estimation (LaggedAlternate)

  • MTR Estimation (Last Good Value)

  • MTR Estimation (Historical Avg)

  • MTR Estimation (Run Ratio)

  • None

TitleResults for “How to create a CRG?”Also Available in