Low flow check validation process

  • Last UpdatedSep 24, 2024
  • 1 minute read

To process Low Flow Check validation

  1. On the Meter Validation tab, located in the Properties for Meter window, enable the MTR Low Flow Check.

  2. Enter the tolerance that would reflect a lower-than-normal flow.

  3. Enter the desired severity. For example, select Valid.

  4. Navigate to the Meter Data Editor window for the meter that had the Low Flow Check enabled.

  5. Locate the row with a corrected volume that is below the set tolerance.

  6. Right-click on the row and select Validate.

  7. Refresh the Meter Data Editor window to make the data quality reflect the severity chosen at the time when the Properties were being configured.

  8. Navigate to the Problem tab on the Meter Data Editor window.

    The validation error is recorded if the severity is set to either Flagged or Invalid. In addition, any new data coming through the system is validated.